Cảm biến Oxy hòa tan quang học
Cảm biến oxy hòa tan hiệu năng cao dành cho các ngành công nghiệp thực phẩm, đồ uống, các ứng dụng vi điện tử, nước tinh khiết
Product
InPro 6860 i |
Easy handling – exceptional performance
• No electrolyte or polarization time
required
• Automatic stability control for superior
measurement stability
• Plug and Measure for fast, error-free
start up
• Pre-calibration in the lab with
iSense
• Supports analog (nA / mA) and digital
(Modbus) direct system integration
• Improved optical technology ensures
high accuracy
|
InPro 6960 i |
Specifically for wort aeration
• Long maintenance intervals thanks to
robust design of sensing element
• One-minute maintenance
• Full ISM functionality for fast start-up and
proactive maintenance
• Customizable internal LED switch-off option
for extended sensing element lifetime
• Hygienic design minimizes contamination
risks
• No electrolyte or polarization time required
|
InPro 6970 i |
Top performance and intelligent diagnostics
• Detection down to 2 ppb in beer and
water
• Fast response time minimizes beer losses
• Sensing element is immune to pressure
shocks and extreme CIP cycle conditions
• Plug and Measure simplifies handling
and ensures fast, flawless start up
• Predictive diagnostics for proactive maintenance
• Customizable internal LED switch-off option
for extended sensing element lifetime
• No electrolyte or polarization time required
|
Specifications
InPro 6860i | Inpro 6960i | Inpro 6970i | |
Measurement principle | optical | ||
Working conditions | |||
Permissible pressure range during measurement |
0.2…6 bar absolute 2.9…87 psi absolute |
0.2…9 bar absolute 2.9…130.5 psi absolute |
0.2…12 bar absolute 2.9…174 psi absolute |
Mechanical pressure resistance |
max. 6 bar absolute max. 87 psi absolute |
12 bar absolute 174 psi absolute |
|
Permissible temperature range during measuring |
0…60°C 32…140°F |
0…40°C 32…104°F |
|
Mechanical temperature resistance of the sensor shaft |
–20...140°C –4…266°F steam-sterilizable autoclavable |
–20...121°C –4…250°F steam-sterilizable |
|
Analog/digital Interface | |||
nA Output: (InPro 6860i /../ nA) | 60 nA at air (25 °C, 1013 mbar) | ||
mA Output: (InPro 6860i /../mA | 4 to 20 mA (active) | ||
Analog temperature Output: | NTC 22 kΩ | ||
Digital Interface: |
Modbus RTU on RS485 – Baudrate: 4800,9600, 19200 (default), 38400, 56600, 115000 – Device address: 1 (default to 32) |
||
Power requirements |
Umin = 19.5 VDC Umax = ≤25 VDC Pmax = 0.75 W ➝ 33 … 40 mA |
||
Sensor performance | |||
Operating range | 8 ppb to 60% O2 saturation | 2 ppb to 2000 ppb | |
Accuracy in aqueous media | ≤ ±[1%+8 ppb] | ≤ ±[1%+2 ppb] | |
Response time at 25°C (77°F) (air › N2 |
t 98% < 20 s, < 70 s (InPro 6860i) | ||
Design features | |||
Temperature compensation | automatic with built-in RTD | ||
Cable connection (digital) | data cable 5 pin,VP8 (InPro6860i) | ||
Wetted O-rings | EPDM FDA approved (other materials on request) | ||
O2 selective membrane material | silicone, PTFE (InPro 6860i) | ||
Wetted sensor parts | s / steel 1.4404, [AISI 316L] PPS with material certificate 3.1 (other materials on request) | ||
Surface roughness of wetted s / steel parts | |||
(EN1320:1996) | Ra <0.4µm [16µin] | ||
Certificates (MaxCert™) | |||
Quality (final inspection certificate) | yes | ||
FDA/USP Class VI | yes | ||
Material certificate 3.1 | yes | ||
Surface finish certificate 2.1 | yes | ||
ATEX Certificate | InPro 6860i/12/ xxx /mA only |
Bản quyền thuộc về CÔNG TY CỔ PHẦN CÔNG NGHỆ HỢP LỰC. Phát triển bởi LYLE SOFTWARE